15 700 02 0319
Operation
17
ATTENTION!
Inadvertent pulse triggering may result in damage of the device
Only trigger pulses when the handpiece is in contact with the intended treat-
ment zone.
• Avoid excessive pressure of the pulse transmitter to the patient's skin! Excessive
pressure is not necessary for the success of the treatment.
CAUTION!
To prevent the risk of swelling, petechiae and haematomas, do
not apply more than 300 – 400 pulses to the same spot during
treatment.
Apply enough coupling gel to ensure that the pulse transmitter glides smoothly over
the patient's skin.
Setting parameters
Treatment should always start at a low intensity level. This also applies when
resuming treatment after an interruption. The pulse energy should be increased
gradually during treatment.
NOTE
The selection of intensity levels is based on the medical opinion of the person
administering the treatment. The maximum intensity level used during treatment
must not cause the patient undue pain under any circumstances.
Coupling the handpiece
NOTE
Clean all parts which come into contact with the patient
before and after
each treatment.
•
Apply a sufficient amount of coupling gel to the patient’s skin in the treatment
area and to the pulse transmitter.
Avoid excessive pressure of the pulse transmitter to the patient‘s skin! Excessive
pressure is not necessary for the success of the treatment.