I. SAFETY NOTES
NOTE1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration-electromagnetic immunity
The MA801 is intended for use in the electromagnetic environment specified below.
The customer or the user of the MA801 should assure that is used in such and environment.
Immunity test
IEC 60601 test level Compliance level
Electromagnetic
environment-guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 KHz to 80 MHz
6 V in ISM bands
between 0,15 MHz
and 80 MHz 80 %
AM at 1 kHz
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Body
Composition Analyzer MA801
including cables, then the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Radiated RF
IEC 61000-4-3
3 V/m
80MHz to 2,7 GHz
Recommended separation
distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800
MHz d = 2,3 √P 800MHz to 2,7
GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.b
Interference may
occur in the vicinity
of equipment
marked with the
following symbol:
3 Vrms
150 KHz to 80 MHz
6 V in ISM bands
between 0,15 MHz
and 80 MHz 80 %
AM at 1 kHz
3 V/m
80MHz to 2,7 GHz