40
Multi
Tracker
Calibration
Calibration can be done for both zero and lab sample concentration.
fig 7
Zero Sample
Press
to open the menu and step down with
. Select ”Calibration”
with
. Step down with
and select ”zero”. Confirm with
. Put the
sensor in clean, deareated water and cover it to prevent the sensor from
being exposed to light and confirm with
.
. When zero calibration is
done, exit with
until one of the two main windows appears.
Conc. sample – Sludge Sample
Press
to open the menu and step down with
. Select ”Calibration”
with
. Step down with
and select ”Sample”. Confirm with
. Put the
senor in a sample with known concentration and stir sample with sensor to
keep solids in suspension. Confirm with
. When complete then exit with
until one of the two main windows appears.
Lab Value
Press
to open the menu and step down with
. Select ”Calibration”
with
. Step down with
and select ”Lab”. Confirm with
. Step up
with
or down with
until the value corresponds with the lab value.
Confirm with
. When all settings are done, exit with
until one of the
two main windows appears.