42
Multi
Tracker
storing the measurement. Step up with
or down with
to reach desired log
position. The postion tag will be shown in the lower margin in the display window. Fig 8
below, shows that the actual profile will be stored at position ”General”
fig 8
It is possible to tag each of the 250 log positions with up to 10 alphanumerical
characters. By entering the log position by tag name, all actual information for the stored
measurement may be retrieved.
Alarm values
At delivery the MultiTracker is preset with the following values; Threshold 1, 500mg/l
(Low suspended solids level) and Threshold 2, 5,000mg/l (High suspended solids level).
Using these values, it is possible to instantly start to measure. To change the settings
please see sections for Threshold 1 and Threshold 2.
Maintenance
The Solido is designed to reduce the manual maintenance to a minimum. All metal parts
are stainless steel (SS 2343/SS316). The enclosure is IP68/NEMA 7. The sensor cable is
a specially manufactured PUR™ with a strong shield and extra heavy wires to withstand
mechanical wear for a long life. The sensor and enclosure cable fittings are high quality
MatchClamp™ to ensure a water proof connection even should the outer shield be
damaged.
Inspection of the sensor
The sensor head should be cleaned if any solids or fouling of the measuring windows
occurs. In order to verify the necessity of cleaning, place the sensor in clean, de-aerated
water and read the display value. The value should not differ more than ±10mg/l from
zero. If the value is off the cleaning is required, a new zero calibration may be
performed. See section
When cleaning the sensor, it is preferred to use the Cerlic Sensor Cleaning (CSC) liquid.
It is also possible to use a soft cloth and water. Pay attention not to scratch the optical
windows.
Trouble shooting
In case of any malfunction that is not possible to correct with a new calibration, please
contact Cerlic or a local Cerlic representative. In case of sending the Solido to Cerlic for
checkup or repair, please make sure to use the form for Return of Material (RMA) prior
to dispatch. The RMA document can be downloaded from the Cerlic web page
. The WEB page also has the actual and correct receiving address.