Changing/installing an Optical cell
Unscrew the electrode holder, release the contact.
Push out the electrode from the back, clean the electrode holder.
Change o-ring (29,87x1,78) apply silicone lubricant, also on the o-ring
around the electrode. Carefully mount the electrode in the holder and reconnect.
Turn the electrode holder anticlockwise to get the cables right. Tighten.
NOTE! Flushing max 1 time/hour, max 2 bar and max 10 sec.
Getting Started
Getting started with the Oxyduo
Start the MultiTracker by pressing the button marked
. To switch OFF/ON the unit,
press the same button. The unit will start automatically. Values for oxygen
concentration are displayed in clear text and real time. If the unit is not active during an
eight minute period then it will automatically be turn off without saving the
measurements.
Save a measurement
To save a measurement, or to perform a new measurement, press
, when doing this,
an option to save
or restart the measurement
will be presented. There is a
short command from the text mode display where it is possible to select position for
storing the measurement. Step up with
or down with
to reach desired log
position. The postion tag will be shown in the lower margin in the display window. Fig 8
below, shows that the actual profile will be stored at position ”General”
fig 8
It is possible to tag each of the 250 log positions with up to 10 alphanumerical
characters. By entering the log position by tag name, all actual information for the stored
measurement may be retrieved.
Alarm values
At delivery the MultiTracker is preset with the following values; Threshold 1, 9mg/l
(High oxygen level) and Threshold 2, 2mg/l (Low oxygen level). Using these values, it is
possible to instantly start to measure. To change the settings, please see Threshold 1
and Threshold 2.
Maintenance
The Oxyduo is designed to reduce the manual maintenance to a minimum. All metal
parts are stainless steel (SS 2343/SS316). The sensor cable is a specially manufactured
PUR™ with a strong shield and extra heavy wires to withstand mechanical wear for a
long life. The sensor and enclosure cable fittings are high quality MatchClamp™ to
ensure a water proof connection even should the outer shield be damaged.