1220C GB 19991004
Basic installation requirements
Before unpacking the unit, make sure you have a darkroom with an adequate
working area and the basic installation requirements, as listed below. Complete
specifications are provided in Appendix A.
Installation requirements
■
Power requirement
Prior to connecting the processor to the mains, check whether the specifica-
tions on the data plate match your local power supply (120V / 230V). The
processor must not be run without earth protection. An earthed power outlet
protected by a fuse (15A / 16A) is required. An earth leakage breaker
(IN = < 30 mA) must be built into the electric circuit.
■
Water requirement
The Cawomat 2000 IR processor does not need a separate water connection.
Water is supplied by a water replenishment bottle, and should be at normal
room temperature (15°C to 25°C / 59°F to 77°F). A convenient sink and water
hose is recommended for filling the water replenishing bottle, for mixing the
chemicals and for cleaning the processor.
An optional fixed water connection can be ordered though, if continuous water
supply is wanted. Refer to page 87 for the appropriate order code.
■
Disposal of waste chemicals and waste water
It is the purchaser’s responsibility to conform to local rules and regulations re-
garding the disposal of waste chemicals. Care should be taken when handling
chemicals at all times. Please observe the hazard instructions provided on the
chemical labels.
■
Ventilation requirement
The Cawomat 2000 IR processor does not require a separate exhaust system
and does not produce excessive external heat. The processor will operate prop-
erly at a room temperature between 15°C and 30°C / 59°F and 86°F, and at a
relative humidity between 30% and 80%.
9
Installation
1
1220C Cawomat 2000 IR RM 07-10-99 14:17 Page 9
Содержание 2000 IR
Страница 1: ...CAWOMAT 2000 IR Reference manual Cawomat 2000 IR 1220C Cawomat 2000 IR RM 07 10 99 14 17 Page p1...
Страница 5: ...Preface Introducing the Cawomat 2000 IR 1220C Cawomat 2000 IR RM 07 10 99 14 17 Page 1...
Страница 6: ...1220C GB 19991004 2 Preface 0 1220C Cawomat 2000 IR RM 07 10 99 14 17 Page 2...
Страница 12: ...1220C GB 19991004 8 Installation 1 1220C Cawomat 2000 IR RM 07 10 99 14 17 Page 8...
Страница 22: ...1220C GB 19991004 18 Installation 1 1220C Cawomat 2000 IR RM 07 10 99 14 18 Page 18...
Страница 24: ...1220C GB 19991004 20 Getting started 2 1220C Cawomat 2000 IR RM 07 10 99 14 18 Page 20...
Страница 38: ...1220C GB 19991004 34 Cleaning 3 1220C Cawomat 2000 IR RM 07 10 99 14 18 Page 34...
Страница 50: ...1220C GB 19991004 46 Troubleshooting 4 1220C Cawomat 2000 IR RM 07 10 99 14 18 Page 46...
Страница 87: ...A Appendix Equipment information sheet 1220C Cawomat 2000 IR RM 07 10 99 14 20 Page 83...
Страница 88: ...1220C GB 19991004 84 Specifications A 1220C Cawomat 2000 IR RM 07 10 99 14 20 Page 84...
Страница 93: ...B Appendix Parts list and exploded views 1220C Cawomat 2000 IR RM 07 10 99 14 20 Page 89...
Страница 94: ...1220C GB 19991004 90 Parts list B 1220C Cawomat 2000 IR RM 07 10 99 14 20 Page 90...
Страница 135: ...C Appendix Index 1220C Cawomat 2000 IR RM 07 10 99 14 27 Page I 1...
Страница 136: ...1220C GB 19991004 I 2 Index C 1220C Cawomat 2000 IR RM 07 10 99 14 27 Page I 2...
Страница 141: ...1220C Cawomat 2000 IR RM 07 10 99 14 27 Page I 7...