Software Communication
3-44
Table 3-7. Error Codes
Error Code
Description
0 (00h)
Error Free Condition.
1 (01h)
Initialization error. This error occurs when the pump fails to initialize.
Check for blockages and loose connections before attempting to
reinitialize. The pump will not accept commands until it has been
successfully initialized. This error can only be cleared by successfully
initializing the pump.
2 (02h)
Invalid Command. This error occurs when an unrecognized command is
issued. Correct the command and operation will continue normally.
3 (03h)
Invalid Operand. This error occurs when an invalid parameter (<n>) is
given with a command. Correct the parameter and pump operation will
continue normally.
4 (04h)
Invalid Command Sequence. This error occurs when the command
structure or communication protocol is incorrect. Review the information
describing the communication protocol then repeat the command sequence.
5 (05h)
Fluid Detection. This error occurs when the sensor board detects fluid,
which is caused by fluid leaking out the back of the valve.
To clear the error:
1
Power off the pump.
2
Remove the valve.
3
Wipe up any detectable fluid on the sensor board using a cotton swab.
You can insert the swab through the valve opening on the front panel,
and gently wipe the circuit.
4
Assuming the valve is leaking fluid, put a new valve on the pump
following the instructions in Chapter 5, “Maintenance.”
5
Reinitialize the pump.
6 (06h)
EEPROM Failure. This error occurs when the EEPROM is faulty. If you
receive this error, please call Cavro Technical Service.
7 (07h)
Device Not Initialized. This error occurs when the pump is not initialized.
To clear the error, initialize the pump.
9 (09h)
Plunger Overload. This error occurs when movement of the syringe
plunger is blocked by excessive back pressure. The pump must be
reinitialized before normal operation can resume. This error can only be
cleared by reinitializing the pump.
10 (0Ah)
Valve Overload. This error occurs when the valve drive loses steps by
blockage or excess back pressure. The pump must be reinitialized before
normal operation can resume. Sending another Valve command
reinitializes the valve and sets it to the correct location. Continual valve
overload errors are an indication the valve should be replaced.
11 (0Bh)
Plunger Move Not Allowed. When the valve is in the bypass or throughput
position, Plunger Movement commands are not allowed.
15 (0Fh)
Command Overflow. This error occurs when the command buffer contains
too many characters. Commands in the buffer must be executed before
more commands can be sent.
Содержание XP 3000
Страница 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Страница 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Страница 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Страница 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Страница 134: ......