XP 3000 Physical Specifications
E-2
Interface
Type
RS-232, RS-485 or CAN
Baud Rate
9600 or 38400 (RS-232 and RS-485 only)
Format
Data Bits: 8
Parity: No
Stop Bit: 1
Half Duplex
Communications
Addressing
Up to 15 pumps can be addressed individually (up to 16 pumps w/
microstep FW)
Communications
Data terminal and OEM protocol (with error recognition)
Firmware
Programmable Ramps
Programmable Backlash Compensation
Programmable Plunger Speeds
Programmable Delays
Programmable Loops
Change Speed on the Fly
Terminate Moves
Diagnostics
Absolute or Relative Positions
Programmable EEPROM
Inputs
Two TTL level inputs with 4.7k pull ups
Outputs
Three outputs, CMOS (HC) level
Environmental
Operating Temperature
(mechanism)
59
°
F (15
°
C) to 104
°
F (40
°
C)
Operating Humidity
(mechanism)
20-95% RH at 104
°
F (40
°
C)
Storage Temperature
-4
°
F (20
°
C) to 149
°
F (65
°
C)
Содержание XP 3000
Страница 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Страница 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Страница 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Страница 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...