Hardware Setup
2-11
C
ONFIGURATION
J
UMPER
JP1
The table below describes the features of JP1.
Jumper
Standard Running Mode
JP1-1
Overload Disable
JP1-2
Reserved
JP1-3
Self-Test and EEPROM
JP1-4
Baud Rate
JP1-1, Plunger Overload Detection
This jumper position allows enabling or disabling of plunger overload detection. There
are two possible settings:
JP1-1 removed
Plunger overload detection enabled (default setting)
JP1-1 installed
Plunger overload detection disabled
JP1-2, Reserved
JP1-3, Self-Test and EEPROM AutoStart
When JP1-3 is in and the address switch is set to position “F,” the self-test diagnostic
program will be activated. For more information on using the address switch and the self-
test, see “Address Switch Settings” in this chapter.
When the address switch is in any position other than “F,” the autostart mode of the
EEPROM can be activated or inactivated.
JP1-3 removed
EEPROM, self-test disabled (default setting)
JP1-3 installed
EEPROM self-test enabled
JP1-4, Baud Rate
This jumper position is used to select the baud rate for the RS-232/RS-485 version of the
XP 3000. There are two baud rates to select from:
JP1-4 removed
9600 baud (default setting) (100K baud for CAN)
JP1-4 installed
38400 baud (125K baud for CAN)
Содержание XP 3000
Страница 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Страница 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Страница 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Страница 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Страница 134: ......