Software Communication
3-17
Example 2. The pump is set to address 6
Pump sends:
Dir
Group
Device
Frame
RTR
Length
1
001
0110
010
0
0000
Host acknowledges:
Dir
Group
Device
Frame
RTR
Length
Node ID
Slave ID
0
001
0000
000
0
0010
0010 0110
0010 0110
Host acknowledges the boot request with:
Dir = 0
Host to slave
Group = 1
Boot request response group
Device = 0
Always 0 in boot response
Note
:
Boot MID is the same for all nodes
Frame = 0
Boot request response frame
Rtr = 0
Always 0
Length = 2
Two data bytes in return message
Node ID
Group ID (2) + Pump
Address (6)
“&”
Hex
26
Slave ID
Same as Node ID
(hex 26)
Hex
26
The pump will save the Node ID to use for message filter Group ID.
CAN H
OST AND
P
UMP
E
XCHANGES
When a slave pump receives a command, finishes a command, encounters an error
condition, or responds to a query, it sends an answer frame to the host using the same
frame type as the command it belongs to. The answer frame format is device dependent.
Generally, it will have the following format:
<MID><DLC><Answer>
Where:
<MID>:
11-bit message identifier. The direction bit is 1. The group number and
the frame type are the same as received. Device is the current device
address.
<DLC>:
4-bit data length code.
<Answer>:
Data bytes block. The first byte of the data block is always the status
byte. It is defined as in Table 3-. The second byte is a null character.
The remaining bytes contain the response in ASCII format. If the reply
consists of more than six bytes, the multi-frame messages are used.
Содержание XP 3000
Страница 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Страница 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Страница 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Страница 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Страница 134: ......