Software Communication
3-28
Table 3-6. Recommended Initialization Forces by Syringe
Parameter
Force
Syringes
0, 3-40
Full
1.0 mL and larger
1
Half
50, 100, 250, 500 µL
1
Z = Z0
2)
2 - 9 are reserved
3)
Z10-Z40 (or Y10-Y40) are initialization speeds which correspond to ([S] commands,
Set Speeds) S10-S40 found in “Set Commands (Velocity and Acceleration)” in this
chapter. These commands can be used to change the standard initialization speeds.
Slower initialization speeds may be useful when working with viscous fluids or
small I.D. (inner diameter) tubing.
I
NITIALIZATION
C
OMMANDS FOR
V
ALVE
T
YPES
3-P
ORT
, 3-P
ORT
D
ISTRIBUTION AND
T-V
ALVE
Z <n> Initialize Plunger (Set Output Valve to Right)
The [Z] command initializes the plunger drive and sets valve output to the right (as
viewed from the front of the pump). The parameters are described below.
Command
Parameter
Description
Z
<n> = 0
Initializes at full plunger force
<n> = 1
Initializes at half plunger force
Y <n> Initialize Plunger (Set Output Valve to Left)
The [Y] command initializes the plunger drive and sets valve output to the left (as viewed
from the front of the pump). The parameters are described below.
Command
Parameter
Description
Y
<n> = 0
Initializes at full plunger force
<n> = 1
Initializes at half plunger force
I
NITIALIZATION
C
OMMANDS FOR
V
ALVELESS OR
B
LOCK
U
NITS
W <
N
> Initialize Plunger (Without Valve)
The [W] command initializes plunger drive for pumps without valves.
Command
Parameter
Description
W
<n> = 0
Initializes at full plunger force
<n> = 1
Initializes at half plunger force
Содержание XP 3000
Страница 21: ...Hardware Setup 2 5 RS 232 CABLING Figure 2 3 RS 232 Multi Pump Cabling nc ...
Страница 22: ...Hardware Setup 2 6 RS 485 CABLING Figure 2 4 RS 485 Multi Pump Cabling RS 485 A RS 485 B ...
Страница 23: ...Hardware Setup 2 7 CAN CABLING Figure 2 5 CAN Multi Pump Cabling ...
Страница 36: ...Hardware Setup 2 20 Figure 2 12 XP 3000 Threaded Mount Holes ...
Страница 134: ......