46 |
P a g e
cascadesciences.com
|
Tel. 503 847-9047
OPERATION
O
VEN
C
OOLDOWNS
The oven chamber is well insulated and requires a significant length of time to cool down while
remaining sealed.
•
6 hours may be required to cool down from 150°C to 100°C.
•
16 hours may be needed for the oven to return to room temperature cooling from 150°C.
•
Backfilling the oven does not significantly increase the rate of cooling.
Introducing free atmosphere into the oven chamber at temperatures above 100°C risks oxidizing
chamber surfaces.
Содержание CVO-2
Страница 1: ...Installation Operation Manual CVO 2 Benchtop Vacuum Ovens...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Страница 20: ...20 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Страница 32: ...32 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Страница 50: ...50 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Страница 52: ......