27
|
P a g e
cascadesciences.com
|
Tel. 503 847-9047
INSTALLATION
S
HELVING
I
NSTALLATION
Heating in a vacuum environment takes place partly through conduction.
Heat is transported from oven elements inside the chamber walls or floor to
the shelves. Install the shelves as described below to ensure proper heat
conduction and temperature measurement.
Never place samples or product on the oven chamber floor. The floor runs
hotter than the shelf temperatures. All oven heating specifications are for
shelving temperatures only.
1. Carefully slide the short shelf into position on the chamber floor, sliding
the clip on the bottom of the shelf onto the oven temperature probe.
•
The oven probe extends from the back wall near the floor of the
chamber.
•
The short shelf must be on the bottom
of the shelf-stack to ensure
the oven controls temperature properly.
2. Place the 2 tall shelves on top of the short shelf.
3. Place the 2 medium shelves on top of the tall shelves.
CVO-2 Shelving Configuration
Temperature Probe Shelf Clip
Bottom Shelf Upside Down
1
2
3
Temperature Probe
Содержание CVO-2
Страница 1: ...Installation Operation Manual CVO 2 Benchtop Vacuum Ovens...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Страница 20: ...20 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Страница 32: ...32 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Страница 50: ...50 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Страница 52: ......