UNIT SPECIFICATIONS
Please refer to the oven data plate for individual electrical specifications.
Technical data specified applies to units with standard equipment at an ambient temperature of 25°C
±3° (77°F ±5.4°) and at nominal voltage. The temperatures specified are determined in accordance
with factory standards respecting the recommended wall clearances of 10% of the height, width, and
depth of the inner chamber. All indications are average values, typical for units produced in the
series. We reserve the right to alter technical specifications at all times.
T
EMPERATURE
P
ERFORMANCE
Operating Range
Fahrenheit
Celsius
A18° to 300
°
F
A10 to 150
°
C
Temperature Uniformity
@ 105°F
@ 200°F
@ 300°F
±1.0°F
±1.5°F
± 2.2°F
@ 40°C
@ 90°C
@ 150°C
±0.5°C
±0.8°C
± 1.2°C
Temperature Stability
@ 105°F
@ 200°F
@ 300°F
±1.0°F
±1.5°F
± 2.2°F
@ 40°C
@ 90°C
@ 150°C
±0.05°C
±0.04°C
± 0.2°C
Cooldown Times
Approximate times with the chamber under vacuum. Cool down rate will be influenced by ambient
conditions.
From 105°F
From 200°F
From 300°F
~255 minutes to 86°F
~595 minutes to 122°F
~595 minutes to 122°F
From 40°C
From 90°C
From 150°C
~255 minutes to 30°C
~595 minutes to 50°C
~595 minutes to 50°C
Содержание CVO-2
Страница 1: ...Installation Operation Manual CVO 2 Benchtop Vacuum Ovens...
Страница 6: ...6 P a g e cascadesciences com Tel 503 847 9047 TABLE OF CONTENTS...
Страница 8: ...8 P a g e cascadesciences com Tel 503 847 9047 CERTIFICATIONS...
Страница 12: ...12 P a g e cascadesciences com Tel 503 847 9047 SPECIFICATIONS...
Страница 20: ...20 P a g e cascadesciences com Tel 503 847 9047 RECEIVING...
Страница 30: ...30 P a g e cascadesciences com Tel 503 847 9047 INSTALLATION...
Страница 32: ...32 P a g e cascadesciences com Tel 503 847 9047 SYMBOLS...
Страница 50: ...50 P a g e cascadesciences com Tel 503 847 9047 MAINTENANCE...
Страница 52: ......