10-2. Error message in the Weighing Mode
Err No
Origin
Solution
Err 01
The weight is too unstable to initialize the scale. Lay the scale on a plat place and turn on the power.
Err 02
The Load cell connection is failed or an error
occurs in A/D conversion part.
Check the load cell connector to see if the polarity of
the signal is reversed.
Err 05
Any key is pressed for a long time or the key
part has problem.
Contact the service center.
Err 08
The ZERO key or TARE key is adjusted not to
be operated under the unstable condition.
Reset the operational condition of the ZERO and
TARE key in the Set Mode F14.
Err 09
The Current weight deviates from the zero range.
Set the operational range of the ZERO key within
2% or 10% of the maximum capacity in the Set Mode
F13.
Err 10
The tare weight exceeds the maximum capacity
of the scale.
Set the tare weight to be smaller than the maximum
capcacity.
Err 13
The zero value deviates from the set range.
Confirm that there is nothing on the weighing
platform.
If nothing exists, execute the Calibration in CAL
Mode.
Err 14
The gross weight is smaller than the Final weight
in Loss-in-batching of the built-in automatic
program Mode.
Set the Gross weight to be bigger than the Final
weight.
Err 15
The Itemcode range is exceeded in the
Command Mode.
Check the ItemCode range.
Err 82
The Error occurs in the A/D conversion part.
Check whether the indicator is properly connected to
the load cell.
Over
The weight on the platform is too heavy to be
measured.
Do not load the item that exceeds the maximum
tolerance.
If the load cell is damaged, it should be replaced.
Содержание NT-580A
Страница 1: ...2 ...
Страница 12: ...13 2 Measure of Appearance 2 1 Measure of Appearance ...
Страница 24: ...25 4 4 Sealing Method Sealing Method of front cal switch ...
Страница 53: ...54 External Control I O Circuit ...
Страница 58: ...59 Normal Batching Diagram Loss in Weight Batching Diagram ...
Страница 59: ...60 UserProgramControlMode NormalBatching ...
Страница 61: ...62 UserProgramControlMode Loss in WeightBatching ...
Страница 63: ...64 AutomaticProgramControlMode NormalBatching ...
Страница 65: ...66 AutomaticProgramControlMode Loss in Weight Batching ...
Страница 85: ...86 MEMO ...
Страница 86: ...87 ...
Страница 87: ...88 ...