20
4. Calibration Mode
4-1. How to enter Calibration Mode
If you open the front cover of the indicator and turn on the DIP S/W2, the Calibration Mode is
started.
At this time,
CAL
message is displayed on VFD and
CAL 1
runs automatically.
If you turn on all of the DIP switches(1,1) after the Calibration, you can return to the Weighing
Mode then.
.
4-2. Available Keys in Calibration Mode
▲
,
▼
Key : Change the set value
▲
Key : increase set value
▼
Key : decrease set value
, Key : Change the digit of set value
▶ ◀
Key : moves one digit to right
▶
Key : moves one digit to left
◀
”* ” Key : Cancel of input value
ENTER Key : The program is moved into the next menu after saving the set value.
4-3. Calibration Menu(CAL1 ~ CAL9)
CAL 1 : Maximum Value Set
CAL 2 : Minimum Division & Dot Position Set
CAL 3 : Weight Calibration
3-1. Set Range of Multi Calibration
3-2. Zero Calibration
3-3. Set Weight
3-4. Span Calibration
CAL 8 : Zero Adjust
CAL 9 : Weight Constant Calibration
Содержание NT-580A
Страница 1: ...2 ...
Страница 12: ...13 2 Measure of Appearance 2 1 Measure of Appearance ...
Страница 24: ...25 4 4 Sealing Method Sealing Method of front cal switch ...
Страница 53: ...54 External Control I O Circuit ...
Страница 58: ...59 Normal Batching Diagram Loss in Weight Batching Diagram ...
Страница 59: ...60 UserProgramControlMode NormalBatching ...
Страница 61: ...62 UserProgramControlMode Loss in WeightBatching ...
Страница 63: ...64 AutomaticProgramControlMode NormalBatching ...
Страница 65: ...66 AutomaticProgramControlMode Loss in Weight Batching ...
Страница 85: ...86 MEMO ...
Страница 86: ...87 ...
Страница 87: ...88 ...