22
CAL 1: The step where the maximum capacity is set.
Function: Maximum capacity set-up (Range: 1~99,999)
Key used
Display
Sub-Display
Description
: Save and/or
move to the next menu
~
: Change the value
: Finish
C= 500
C=2000
CAL 1
CAPA
500 kg
2000 kg
Note 1. The maximum capacity refers to the maximum value of weight the scale can measure.
CAL 2: The step where the division (the minimum display unit)
is set.
Function: Minimum division set-up (range: 0.001~50)
Key used
Display
Sub-Display
Description
: Save and/or
move to the next menu
~
: Change the value
: Enter a
decimal(.)
d= 1
d= 0.2
d= 0.05
d=0.001
CAL2
dIVISIon
1 kg
0.2 kg
0.05 kg
0.001 kg
Note 1. The minimum division refers to the value of 1 division.
Note 2. The minimum division can be set from 0.001 ~ 50. The minimum division in single
digit is available as 1, 2, and 5. In double digits, it is available as 10, 20, and 50. Same
goes for numbers with decimals.
Note 3. Do not set the value of 'one division/the maximum capacity' as more than '1/ 50,000',
respectively. If it is over 1/50,000, an error message will be displayed. 1/50,000 is the
maximum value.
Содержание CI-502A
Страница 1: ... 2 ...
Страница 9: ... 10 External Appearance and Dimensions External Appearance CI 502A CI 505A ...
Страница 10: ... 11 Dimensions of the Product CI 502A CI 505A ...
Страница 11: ... 12 Front Panel Descriptions CI 502A CI 505A ...
Страница 27: ... 28 9 th Step Item MovebacktotheWeighingMode ...
Страница 67: ... 68 Weight Data 8 byte a 13 5kg 1 3 5 b 135kg 1 3 5 c 135kg 1 3 5 ...
Страница 74: ... 75 Sealing Method Sealing 1 Sealing the Indicator Sealing the Load Cell Connector ...
Страница 83: ... 84 MEMO ...
Страница 84: ... 85 MEMO ...
Страница 85: ... 86 MEMO ...
Страница 86: ... 87 MEMO ...
Страница 87: ... 88 ...
Страница 88: ... 89 ...