21
(2) Calibration Menu (CAL 1 ~ CAL 7)
P 1 : Multi-calibration range set-up
CAL 1 : Maximum Capacity Set-up
CAL 2 : Minimum Division Set-up
CAL 3 : Zero Calibration
CAL 4 : Setting the Balance Weight (Setting Weight) during Span Calibration
CAL 5 : Span Calibration
CAL 6 : Weigh Constant Calibration
CAL 7 : Equivalence input mode(DIRECT Loadcell input mode)
P 1 : Designate the Number of Calibration
Function: Multi-calibration range set-up
Display
Sub-Display
Description
Set Value
(1~4)
P 1
MuLtI CAL
__ ___ ___1-4
Calibrate the maximum weight to
1~4 value range
* If the load cell’s actual curve is a straight line, set the value to 1.
* If the load cell’s actual curve is not a straight line as shown in the figure below,
this function is used to set multi-points to compensate for the load cell’s output.
* How to enter the Calibration mode
Upon specifying the number of calibration steps, the display for entry into calibration will
be displayed.
* Screen Display: 1-7 displayed
You can select and designate according to steps.
(zero span)
Zero and Span points to interpolate weight from Load Cell
Multi-point
Multi-point
mV/V
Real curve
weight
(full span)
Multi-point
Ideal
Multi-point
Multi-point
Содержание CI-502A
Страница 1: ... 2 ...
Страница 9: ... 10 External Appearance and Dimensions External Appearance CI 502A CI 505A ...
Страница 10: ... 11 Dimensions of the Product CI 502A CI 505A ...
Страница 11: ... 12 Front Panel Descriptions CI 502A CI 505A ...
Страница 27: ... 28 9 th Step Item MovebacktotheWeighingMode ...
Страница 67: ... 68 Weight Data 8 byte a 13 5kg 1 3 5 b 135kg 1 3 5 c 135kg 1 3 5 ...
Страница 74: ... 75 Sealing Method Sealing 1 Sealing the Indicator Sealing the Load Cell Connector ...
Страница 83: ... 84 MEMO ...
Страница 84: ... 85 MEMO ...
Страница 85: ... 86 MEMO ...
Страница 86: ... 87 MEMO ...
Страница 87: ... 88 ...
Страница 88: ... 89 ...