CuroCell® IQ Cirrus
11
5
Operation CuroCell® IQ
Operation CuroCell® IQ (how to get started)
1. Set the On/Off switch on the side of the control unit to position 1 (On). See figure 1.
2. The mattress starts to inflate. This takes about 20 minutes depending on the size of the
mattress. While the mattress is inflating, the “silent information signal” and “information
signal” diodes light up in orange. When these diodes goes out, the user can be placed on the
mattress.
3. The control unit sets the inner pressure of the mattress according to the weight, length, and
position of the user. This takes about 20 minutes. During this time, the diode above the “check
symbol” flashes. When the diode stops flashing, the mattress is set and the inner pressure has
adjusted according to the user.
4. Perform a hand check to ensure that the settings are correct.
Figure 1. On/Off switch.
Note!
•
During the automatic setting, try to avoid
larger movements on the mattress.
Otherwise, you will get a notification on that
the desired value could not be reached
within the time limit and the weighing must
start over.
•
The mattress must be inflated before the
user can lie on the mattress.
•
Once automatic setting is complete, the
control unit switches to a basic setting of
Pulsating mode (when used for the first time)
or to the previous setting.