CuroCell® A4 Cirrus
10
Notification
Description and troubleshooting
1
High temperature. Valves and compressors are turned off. If
the control unit is in direct sunlight, relocate it.
2
Default settings are not completed.
3
Incorrect input voltage. Make sure that the correct power
supply is used, otherwise contact the technical support.
4
Low pressure. Secure the CPR, mattress, air tubes and air
filter.
5
Automatic setting failure. The correct pressure has not been
reached within the time limit.
6
Leakage under the automatic setting period. The mattress is
leaking too much for the weighing to be accomplished. Control
the mattress and the air connections.
7
High pressure. The pressure cannot be reduced to the desired
value within the time limit.
8
The automatic setting has been restarted too many times
during the automatic setting period.
9
The mattress control parameters have not been read. Connect
the CPR.
10
The mattress control parameters have been changed during
the use. Restart the system.
11
Leakage in one of the sections. Secure the CPR, mattress and
connection tubes. Information to service technicians: this
notification shows a leakage in the blue air cell section. More
information in the Servicemanual for CuroCell® A4.
12
Leakage in one of the sections. Secure the CPR, mattress and
connection tubes. Information to service technicians: this
notification shows a leakage in the green air cell section. More
information in the Servicemanual for CuroCell® A4.
13
Leakage in one of the sections. Secure the CPR, mattress and
connection tubes. Information to service technicians: this
notification shows a leakage in the red air cell section. More
information in the Servicemanual for CuroCell® A4.