CuroCell® IQ Cirrus
13
5.2
Program
Pulsating mode with a cycle period of 10
minutes is preset.
5.3
Maximum pressure (caring mode)
With this function, the entire
mattress Is inflated and provides
firmed support. This function reverts
to the previous setting after 20 minutes. The
function should be used when caring the user,
shifting the user’s position or moving the user
in or out of bed.
5.4
Hand check (function control)
Hand check is performed to ensure that the
mattress system works properly. Hand check
should be performed regularly; for CuroCell® IQ
Cirrus, we recommend once per work shift as
well as after installation of the system.
1. Open the cover and insert a hand, with the
palm facing up, between the mattress and the
bed base. The hand shall be inserted beneath
the user’s sacrum (center of mattress).
2. Ensure there is a gap between the user and
the bed base so that the user does not ‘bottom
out’.
3. If you can feel the user’s sacrum resting in
the palm of your hand, the gap is too small. See
section 8 ‘Troubleshooting’.
5.5
CPR (Cardiopulmonary resuscitation)
See 3.2
5.6
Transport function
See 3.3
5.7
Pack&Go®
The product may be packed as follows:
Ensure that no-one is lying on the
mattress.
On the control panel, press the
lock/unlock button.
Press the Pack&Go® button and hold it
down for 2 seconds
The Pack&Go® diode will flash during deflation.
The mattress will empty of air and be ready to
be simply folded together in about 20 minutes.
The control unit gives an audio signal once
deflation is complete.
Carefully fold the mattress in half, place the
control unit between the folds of the mattress
and place the system in a transportation bag
(accessory) or equivalent for protective storage.
Ensure that the power supply is packed
complete.
5.8
Restart
See 3.4
5.9
Power failure
See 3.6
5.10
Notifications
Different notifications exist based on
how serious the warning is. With a
malfunction or an error, a notification
will be given by way of a flashing
warning triangle. To mute the warning
signal, press the mute button. The
notification code is
shown on the four
different diodes above
the wrench symbol.
5.11
Maximum pressure notifications.
See 3.5
Note!
•
Make sure that the mattress system is filled,
which is shown by a green light from the
diode, before performing a hand check.