CuroCell® A4 Cirrus
9
4.6
Hand check (function control)
Hand check is performed to ensure that the
mattress system works properly. Hand check
should be performed regularly; for CuroCell® A4
Cirrus, we recommend once per work shift as
well as after installation of the system.
1. Open the cover and insert a hand, with the
palm facing up, between the mattress and the
bed base. The hand shall be inserted beneath the
user’s sacrum (center of mattress).
2. Ensure there is a gap between the user and
the bed base so that the user does not ‘bottom
out’.
3. If you can feel the user’s sacrum resting in the
palm of your hand, the gap is too small. See
section 8 ‘Troubleshooting’.
4.7
Sitting positioning in bed
See 3.1
4.8
CPR (Cardiopulmonary resuscitation)
See 3.2
4.9
Transport function
See 3.3
4.10
Pack&Go® function
The product may be packed as follows:
Ensure that no-one is lying on the mattress.
•
On the control panel, press the
lock/unlock button.
•
Press the Pack&Go® button and hold it
down for 2 seconds
The Pack&Go® diode will flash during deflation.
The mattress will empty of air and be ready to be
simply folded together in about 20 minutes. The
control unit gives an audio signal once deflation
is complete.
Carefully fold the mattress in half, place the
control unit between the folds of the mattress
and place the system in a transportation bag
(accessory) or equivalent for protective
storage. Ensure that the power supply is packed
complete.
4.11
Restart
See 3.4
4.12
Power failure
See 3.6
4.13
Notifications
Different notifications exist based on
how serious the warning is. With a
malfunction or an error, a notification
will be given by way of a flashing warning
triangle. To mute the warning signal, press the
mute button.
When a notification
occurs, the current cycle
time diode will turn off
and a notification code is shown on the four
different cycle time diodes (10, 15, 20, 25). To
read the cycle time during the error notification,
unlock the control panel.
4.14
Maximum pressure notifications
See 3.5
4.15
Table of notifications
Information about each notification is shown in
the notification table: Notifications from 1-10 are
both audible and visual. The error code will be
displayed until the error has been rectified. If the
mute button is pressed the audible warning will
cease for a period of 5 minutes and will return
until the error has been rectified.
The notifications 11-13 have no audible alarm.
The error code is shown until the system is
restarted.
Note!
•
Make sure that the mattress system is
filled, which is shown by a green light from
the diode, before performing a hand check.