27
“Delete”
“Manual” files can be deleted using the delete function on the Look in ECG
archive menu.
“Automatic” archive files are deleted automatically after successful
transmission to a PC with the ArchPC function.
“Auto-transmission last ECG”
“Yes”: last stored ECG will be automatically sent
“No”: no auto-transmission
“Emergency only”: only last stored ECG acquired in Emergency mode will be
automatically sent .
HRV
analysis
The following can be set: duration of the test (1 to 5 minutes) and reference lead for
analysis.
ECG
Autotimer
The following can be set: number of recordings (intervals); duration of intervals (i.e. time
between recordings); print delay (timed prints and manually activated prints between two
intervals are referred to a sample acquired 10 seconds earlier); number of leads being
printed: twelve (in format set), three or six (selectable).
Arrhythmia
The user can set:
-
the RR advance (in percentage);
-
the RR delay (in percentage);
-
the Print Advance: how many seconds of signal to be printed as reference
normal ECG before the first event (min 2 sec. max 10 sec.);
-
the Print Delay: how many seconds of normal ECG after the last event (min 2
sec. max 10 sec.);
-
Arrhythmia monitor: enable the printout of the abnormal ECG during the test.;
-
Event Print: enable the abnormal ECG printout during the test;
-
ARR Leads: select lead full disclosure report.
Print Delay
It could be possible enable/disable 10 s. print delay, referred to a real time event.