Lead
Sequence
Leads sequence. The sequence of leads printed can be selected. As well as modifying
the order in which the leads are printed out, the selection also influences the rhythm lead
selection menu.
Three options are available:
1. Standard;
2. Cabrera.
3. Personalize: a special sequence can be set.
Analysis
configuration
The type of processing of the ECG trace acquired in automatic mode can be selected.
The choice influences the type of document printed.
Two configuration menus are available, linked to the type of automatic processing
available: "ECG Parameters" and "ECG Interpretation". Processing may also be
disabled by selecting "None".
Inside the single menus
to enable/disable a section.
-
"ECG Parameters".
"Summary" represents the minimum report and cannot be de-selected. The
following are reported: date and time of recording, patient data, note field,
main ECG parameters calculated (heart rate; rhythm type; P, QT, QTc, PR,
QRS and QTr wave amplitudes; frontal vectors; axes).
“ST Measurements” printout of table of ST depression values on the twelve
leads.
“Templates”: printout of templates relative to the twelve leads.
"ECG
Interpretation".
“Summary” represents the minimum report and cannot be de-selected. The
following are reported: date and time of recording, patient data, if inserted,
note field, ECG parameters calculated, frontal vector, indication of normality.
“Rhythm Analysis”: print rhythm strip and diagnosis.
“Interpretation” processing and printing out of ECG interpretation. In
particular: atrial diagnosis, repolarisation disorders, atrial blocks, QRS-T
evaluation.
“Parameters” print complete ECG parameters table.
“Templates”: printout of templates relative to the twelve leads.
Configure
Copy
The "Copy" key can be configured.
1. “Complete”: the last automatic recording is printed out in full.
2. "ECG only": only ECG of last automatic recording is printed out.
3. “Analysis only”: only analysis of last automatic recording is printed out (if available).
Archive
management
ECG traces can be stored in different ways and memory status can be checked. When
operating in “Paper Saving” mode, the saving function is always automatic.
“Save”
“Automatic”: traces are saved automatically when the scan is completed,
with no action by the operator. The operation is indicated on the display by
messages such as: “Saving...” and “ECG saved”.
“Upon request”: at the end of recording, the patient’s file is shown, where the
option “Save&Exit” is available. Press
to proceed. If the test does not
have to be saved, select “”Confirm&Exit”. A copy of the test will still remain
available until a new recording is made or the device is switched off.
“None” saving disabled.