In “Automatic” mode: the twelve ECG leads are printed by groups 2.5
sec per each lead (3x4) plus 10 sec for rhythm leads. Additional pages
are available if diagnosis aids (e.g. Interpretation) are used.
The selected format will be applied to all manual, automatic and autotimed
twelve lead recordings.
Speed of recording on the display and on paper
To select the scrolling speed of the displayed signal:
; the corresponding value is displayed. The options available
are: 12.5 mm/s, 25 mm/s and 50 mm/s.
To select the printing speed:
; the corresponding value is displayed. The options
available are: 5mm/s, 25 mm/s and 50 mm/s.
During printing, the paper transport speed is indicated on the information
line.
Sensitivity of recording on the display and on paper
To select the sensitivity of recording on the display:
; the corresponding value is displayed. The options available
are:
,
5 mm/mV, 10 mm/mV and 20 mm/mV.
To select the sensitivity of recording on printing:
;
the corresponding value is displayed. The options
available are:
,
5 mm/mV, 10 mm/mV and 20 mm/mV.
During printing, the recording sensitivity is indicated on the information line.
Note:
selecting
the sensitivity is set automatically by the device in a way that
optimises the recording over the entire width of the paper. In this case, a sensitivity
of 2.5 mm/mV may be used. This option is recommended for six 3+R lead printing.
To find out how automatic sensitivity is determined, see "Technical Specifications".
Recording filter
If necessary, it is possible to use filters capable of improving the legibility of