![Cardioline ar1200view Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/cardioline/ar1200view/ar1200view_user-manual_2556439043.webp)
43
The display shows the time from the test start.
The test is
automatically terminated at the end of the scheduled time and the
final report is printed out.
to scroll the leads displayed during the test (based on the
electrodes connected).
to interrupt (stop).
No report is printed and the recorded
data are cancelled.
to print copy of the report at the end of the test.
5.9
Recording in “Arrhythmia mode”
Warning
: To obtain a correct recording in “Arrhythmia mode” mode, ensure that the
battery and paper load are enough to perform the selected acquisition.
A recording in Arrhythmia mode allow to analyze in real time the ECG signal
for a predicted period of time in order to detect potential abnormalities of
the Rhythm in the time domain.
During the test if an abnormal rhythm is detected a printout of the events
can be obtained in continuous until the rhythm become normal again (the
length of the printout depends on the setup). At the end of the test, or after
every 5 minutes, a complete report is printed: full disclosure of the
reference ECG lead (selected by the menu) with marker identification of the
abnormal beats (*), patient data summary, table of the computed
parameters.
Note: connect only the electrodes related to the reference lead in order to increase
the patient comfort.
After selecting “Arrhythmia mode”:
To start the recording, the message “Paper, Battery!” will
appear. If the signal has not yet been initialised, the message “Wait …” will
be displayed.
The patient data entry procedure now starts (see “Patient data entry”).
The display shows the time from the test start. The test is automatically
terminated at the end of the default time (5 minutes) and the final report is
printed out (in any case every 5 minutes pressing the key “stop”).