ECG Settings
70-00082-01 A
Atria Electrocardiograph Operating Instructions
4-5
Sequence
Configure automatic functions.
ECG
Menu Items:
Sequence
Description
Selections
Wait for good
data
When enabled (ON), the system delays the ECG reports
until the system has collected ten seconds of low noise
waveform data (all electrodes must be connected). To
force an immediate report, press
[ECG]
again when the
system displays
Waiting for Good Data
. The report
includes “Warning: Data quality may affect interpretation.”
When disabled (OFF), the ECG report starts after 10
seconds of ECG waveform data, regardless of quality. No
warning message is included.
NOTE:
The ECG report does not include waveform data
for disconnected electrodes. Leads with disconnected
electrodes are not used in the ECG analysis.
ON
OFF
Auto Print
Sets ECG report print options (for thermal, plain paper, or
both).
•
ON
—automatically print
•
Prompt
—prompt before printing
•
OFF
—automatically bypass printing
ON
Prompt
OFF
Auto Save
Sets ECG reports save options.
•
ON
—automatically save
•
Prompt
—prompt before saving
ON
Prompt
Auto Send
Requires the Communications option.
See “Communications Setup” on pg. 5-1 and “Setup EMR
Communications” on pg. 5-12.
•
ON
—automatically send
•
Prompt
—prompt before sending
•
OFF
—do not automatically send
ON
Prompt
OFF
Auto Email
Requires the Communications option. Also requires, an
SMTP server address, and an email address.
See “Communications Setup” on pg. 5-1 and “Setup
Email” on pg. 5-10.
NOTE:
By default, automatic email uses the first email
address.
•
ON
—automatically email
•
Prompt
—prompt before sending email
•
OFF
—do not automatically send email
ON
Prompt
OFF
Содержание Atria 3100
Страница 8: ...viii Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 22: ...2 Equipment Setup 2 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 34: ...3 User Interface 3 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 72: ...6 Patient Preparation 6 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 82: ...7 Acquiring ECG Reports 7 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 94: ...8 Patient Records 8 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 98: ...9 Sample Printouts 9 4 Atria Electrocardiograph Operating Instructions 70 00082 01 A Setup Report Continued...
Страница 102: ...9 Sample Printouts 9 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A Measurement Matrix...
Страница 104: ...9 Sample Printouts 9 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 112: ...10 Troubleshooting 10 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 128: ...11 Maintenance and Service 11 16 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 134: ...4 Atria Electrocardiograph Operating Instructions 70 00082 01 A...