Acquiring an ECG Using Pediatric Lead Placements
70-00082-01 A
Atria Electrocardiograph Operating Instructions
7-3
Acquiring an ECG Using Pediatric Lead Placements
In order to use pediatric lead placements, the V3 placement feature must
be enabled. See “V3 Placement” on pg. 4-12. To configure the Atria to
always acquire ECGs using pediatric lead refer to “ECG Settings” on
pg. 4-2.
1. Prepare the pediatric patient, refer to:
• Chapter 6, “Patient Preparation.”
• “Pediatric Chest Lead Placement” on pg. 6-8.
2. Follow steps for “Acquiring ECG Reports” on pg. 7-1.
• For
V3 Placement
, select
V4R
.
NOTE:
In the
Edit Patient Data
field, change the
Age Format
to months and enter the patient’s age in months, or enter the
Date of Birth
, and the Atria calculates the age automatically.
3. Verify the patient’s heart rate displays on the Home screen. If heart
rate does not display, view problem leads on the display. Replace faulty
electrodes as necessary.
4. Enter patient demographics.
NOTE:
To acquire a stat ECG, press
[ECG]
and then press
[Select]
or
[ECG]
to select
Acquire STAT ECG for a new patient
.
5. Press
[ECG]
.
6. Use the navigation buttons to display
Acquire ECG, current patient data
then press
[Select]
.
ECG Reports
The Atria automatically includes a formatted 12-lead report with
demographics in any ECG printout. Additionally, an Atria with the optional
interpretation feature automatically prints the interpretive analysis on the
report (this feature can be disabled).
The Atria can also automatically print the following pages with the ECG:
• Rhythm page
• Copies of ECG, with or without analysis
• Median complex report (interpretive Atria only)
For more information on enabling these reports, or to disable automatic
printing, see “Report Format” on pg. 4-2.
Содержание Atria 3100
Страница 8: ...viii Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 22: ...2 Equipment Setup 2 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 34: ...3 User Interface 3 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 72: ...6 Patient Preparation 6 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 82: ...7 Acquiring ECG Reports 7 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 94: ...8 Patient Records 8 12 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 98: ...9 Sample Printouts 9 4 Atria Electrocardiograph Operating Instructions 70 00082 01 A Setup Report Continued...
Страница 102: ...9 Sample Printouts 9 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A Measurement Matrix...
Страница 104: ...9 Sample Printouts 9 10 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 112: ...10 Troubleshooting 10 8 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 128: ...11 Maintenance and Service 11 16 Atria Electrocardiograph Operating Instructions 70 00082 01 A...
Страница 134: ...4 Atria Electrocardiograph Operating Instructions 70 00082 01 A...