
CAPINTEC, INC.
CRC
®
-25R
ADDING A NUCLIDE
The CRC
®
-25R contains nuclide data (name, half-life, calibration number) for over 80
nuclides.
The user may add up to 10 nuclides. The name, half-life and calibration number will be input.
Press
MENU
from the Measurement Screen.
Select
SETUP
. The
Setup Menu
appears.
Select
OTHER
. Input the password (last 3 digits of the Readout serial number) and press
ENTER
. Figure 6-1 Other Menu appears.
From the
Other Menu
select
NUCLIDES
. The
Nuclides Menu
appears.
1. Add Nuclide
2. Del Nuclide
3. New Cal Num
Figure 6-24 Nuclides Menu
Select
ADD NUCLIDE
to add nuclide data.
Entering Nuclide Name
TYPE IN
NUCLIDE NAME
_
Figure 6-25 Entering Nuclide Name
Use the alphanumeric keys to input the nuclide name. The first 2 characters will be displayed
as letters (press the
SPC
key for a blank if there is only 1 character in the name) and the
next 3 characters as numbers. If the nuclide is in a metastable state, use the
Tc99m
key on
the left of the keyboard for the “m”.
6 - 14
CHAMBER INITIALIZATION
March 15
Содержание CRC-25R
Страница 2: ......
Страница 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 227: ......
Страница 228: ......