CAPINTEC, INC.
CRC
®
-25R
CONSTANCY TEST
Co57
OK? Y OR N
Figure 6-16 Constancy Source Verification
If the displayed source is correct, press
YES
. Figure 6-5 Test Source Menu will re-appear.
If the displayed source is not correct, press
NO
. Figure 6-17 Constancy Test Source
Selection will appear. (In the example below, Co57 and Co60 are set to use as Daily
Sources.)
CONSTANCY TEST
1. Co57
2. Co60
Figure 6-17 Constancy Test Source Selection
Choose the Daily Test Source to be used for the Constancy Test. Figure 6-16 Constancy
Source Verification screen will re-appear.
MOLY ASSAY SETUP
Moly Assay can be performed using a CAPMAC or CANISTER.
When Moly Assay is chosen from the
Test Menu
(reference CHAPTER 10: MEASUREMENT
PROCEDURES, SECTION: MOLY ASSAY), Figure 6-18 Moly Assay Method Selection the
Moly Assay Method Selection screen will appear. However, if only one method is always
used (either a CAPMAC or a CANISTER), the Moly Assay Method Selection screen can be
omitted by selecting either CAPMAC or CANISTER from the Moly Setup in the
Other Menu
.
6 - 10
CHAMBER INITIALIZATION
March 15
Содержание CRC-25R
Страница 2: ......
Страница 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 227: ......
Страница 228: ......