CAPINTEC, INC
CRC
®
-25R
Single Strip Test
The Single Strip Test consists of 2 measurements: the top of the strip and then the bottom of
the strip.
Select
1 STRIP
from Figure 11-32 QC Test Selection Menu. Figure 11-33 1 Strip
Measurement Message – Top will appear.
Place TOP of Strip
in Chamber
Any Key to Continue
Figure 11-33 1 Strip Measurement Message – Top
Place the Top of Strip in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-34 QC Test Measurement Screen will appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 11-34 QC Test Measurement Screen
Press
ENTER
to accept the measurement. Figure 11-35 1 Strip Measurement Message –
Bottom screen will appear.
Place BOTTOM of Strip
in Chamber
Any Key to Continue
Figure 11-35 1 Strip Measurement Message – Bottom
August 07
ENHANCED TESTS
11 - 15