CAPINTEC, INC
CRC
®
-25R
Press any key (except
HOME
) to continue to the next Results screen.
CALICHECK TEST
MEAN: n.nnnmCi
# RESULT % VAR
5 x.xxxmCi x.xx
6 x.xxxmCi x.xx
7 x.xxxmCi x.xx
8 x.xxxmCi x.xx
Any Key to Continue
Figure 11-31 Calicheck – 2
nd
Test Results Screen
If more than 8 tests have been selected in the test setup, a 3
rd
results screen will appear
showing the values for the 9
th
through last test.
Press any key (except
HOME
) to continue. If a printer is attached to the system, a Calicheck
Linearity Test report will print. The report will show the result and the variation from mean for
each tube.
The prompt to print again will appear. To print the test results again, press
YES
. To return to
Figure 11-1 Tests Menu, press
NO
.
QUALITY CONTROL (QC)
Radiochromatography Quality Control tests can be performed using four methods.
Press
TEST
from the Measurement Screen. Figure 11-1 Tests Menu will appear.
Select
. Figure 11-2 Enhanced Tests Menu will appear.
Select
QC
. Figure 11-32 QC Test Selection Menu will appear.
QC TESTS
1. 1 Strip
2. 2 Strips
3. HMPAO
4. MAG3
Figure 11-32 QC Test Selection Menu
11 - 14
ENHANCED TESTS
August 07
Содержание CRC-25R
Страница 2: ......
Страница 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 227: ......
Страница 228: ......