![Capintec CRC-25R Скачать руководство пользователя страница 139](http://html.mh-extra.com/html/capintec/crc-25r/crc-25r_owners-manual_494176139.webp)
CAPINTEC, INC
CRC
®
-25R
Place TOP of
Strip A in Chamber
Any Key to Continue
Figure 11-38 2 Strips Measurement Message – Top A
Place the Top of Strip A in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-39 QC Test Measurement Screen will appear.
Tc99m
6.27mCi
ENTER to Accept
Figure 11-39 QC Test Measurement Screen
Press
ENTER
to accept the measurement. Figure 11-40 2 Strips Measurement Message –
Bottom A screen will appear.
Place BOTTOM of
Strip A in Chamber
Any Key to Continue
Figure 11-40 2 Strips Measurement Message – Bottom A
Place the Bottom of Strip A in the Chamber and press any key (except
HOME
) to perform the
measurement. Figure 11-39 QC Test Measurement Screen will appear.
Press
ENTER
to accept the measurement. Figure 11-41 2 Strips Measurement Message –
Top B screen will appear.
August 07
ENHANCED TESTS
11 - 17
Содержание CRC-25R
Страница 2: ......
Страница 12: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 26: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 58: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 104: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 122: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 192: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 208: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 226: ...CAPINTEC INC CRC 25R This page intentionally left blank ...
Страница 227: ......
Страница 228: ......