CAPINTEC, INC.
CRC
®
-25R
CHAPTER 6
CHAMBER INITIALIZATION
GENERAL
This section describes initialization and parameter setup for the Chamber.
All of these functions are accessed via the Main Menu.
USER KEY ASSIGNMENT
The
USER
Keys (U1, U2, U3, U4, U5) are a quick way to select a nuclide. A nuclide may be
assigned to each key.
To assign keys, press
MENU
from the Measurement Screen.
Select
SETUP
. The
Setup Menu
appears.
Select
OTHER
. Input the password (last 3 digits of the Readout serial number) and press
ENTER
. Figure 6-1 Other Menu appears.
1. User Keys
2. Sources
3. Moly Setup
4. Nuclides
5. Linearity
6. Remote / PC
Figure 6-1 Other Menu
From the
Other Menu,
select
USER KEYS
.
March 15
CHAMBER INITIALIZATION
6 - 1