Technical Specifications
Planning Guide
0180-095-485
OMNERA™400T
Page 7-2
Classification
Classification according to IEC 60601-1
Speed
X-ray
Leakage Radiation
Class
Class I equipment. All dead metal parts of
the equipment are electrical connected to
protective earth.
Applied part
Type B
Protection against ingress of wter
IPXO
Mode of operation
Intermediate use 20%, max 1 min ON / 4
min OFF
Use of anesthetics mixtures
The equipment if not suitable for use in the
presence of flammable anesthetic mixtures
with air, oxygen or nitrous oxide.
Low speed
Maximum speed
Z movement
40 mm/s
150 mm/s
Loading factors
150 kV, 18 mAs, 19 times/hour
G292
0.064 mGy/hour