Component Specification
OMNERA™400T
0180-095-485
Planning Guide
Page 6-9
•
Vertical travel speed = up to 40 mm/sec
•
Table top weight = 47 kg
•
Table weight = 150 kg
•
Table weight = 171 kg (including the imaging unit of approximately 21 kg)
Movements Table Top
•
Movement range, lateral direction (Y) = 300 mm (±150 mm from center).
•
Movement range, longitudinal direction (X) = 0-1200 mm (±600 mm from center).
Movements Detector Holder
•
Movement range, longitudinal direction = up to 800 mm .
Figure 6-8
Cable Outlet
All cables are led through a cable outlet at the bottom of the head end foot plate. It is recom-
mended to lead the cables in a cable duct from foot plate to the closest wall, however a cable
run/channel may also be used. The cables must be enclosed.