![Canon OMNERA 400T Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/canon/omnera-400t/omnera-400t_planning-manual_4018787071.webp)
Component Specification
OMNERA™400T
0180-095-485
Planning Guide
Page 6-3
Movements
Ceiling Wagon Movements
The manual movements of the ceiling wagon is defined by the dimensions of the ceiling wagon
itself and of which rail configuration that is selected. The movement range is also limited by the
mechanical end stops that are placed on the ends of each rail.
Direction
X-rail (mm) Y-rail (mm)
Stroke A
(mm)
Stroke B
(mm)
Standard
3000
4000
3113
2352
Installation cube/Low ceiling
2748
3748
2820
1945
B
A
Y
X