Environment Specifications
Planning Guide
0180-095-485
OMNERA™400T
Page 2-14
Cable Duct Specifications
All cables shall be enclosed in cable channels, cable ducts or conduits according to regulations
(IEC 60601-1).
The following notes shall be considered;
•
Avoid cross-overs and forming of cable coils.
•
Use appropriate dimensions.
•
Consider the bending radius of the cables.
•
If using conduits, consider that the connectors of the high voltage cables should be
able to be pulled through.
•
Cable channels or cable ducts shall be used for on-the-floor installations.