Layout 7 is designed for very small rooms with a low ceiling height (>=2500). In this layout the
Y rails of low ceiling rail type are mounted directly to the ceiling. To minimize the required room
space, the table top could be fixed in the smallest room (in longitudinal direction) and the sys-
tem cabinet is not placed within the room.
This layout enables the same possibilities as the other options but with some restrictions and
limitations.If the table top is fixed, a full patient coverage is not possible without moving the pa-
tient.
Lateral projections from behind the Table are also restricted due to the limited space available.
Focal spot coverage, Beta pos B
Orientation point
Focal spot coverage, Beta pos A
X
Y
659
1945
3
1
1
3
3000
4
0
0
0
Pos A
B
a
c
k
F
ro
n
t
Head
Orientation point
Current Wall stand setup: Tilt right loaded
2100
(
*2
y
)
Pos B
125
4
9
2
Pos D
250 (*1)
750 (*1)
2
5
0
(
*1
)
125 (*3)
1100 (*2x)
8
0
(
*3
)
S
ID
>
2
7
0
0
2
5
0
(
*1
)
3
3
0
(
*3
)
2
0
0
2930
4
0
0
0
Содержание OMNERA 400T
Страница 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Страница 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Страница 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Страница 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Страница 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Страница 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Страница 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Страница 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Страница 89: ...Intentionally left blank...
Страница 90: ...Canon CANON MEDICAL SYSTEMS USA INC...