Room Layout
Planning Guide
0180-095-485
OMNERA™400T
Page 5-18
System Connections
Cables
The cables from the Ceiling suspended unit have two configurations; 16 m or 24 m. The cable
configuration is selected when the System is ordered. The length of the cables is measured
between the generator and the X-ray tube.
The lengths, displayed in the column “Length from outlet” are excluding the cable length need-
ed inside the system cabinet, i.e. the 16 m configuration has 6 m of cables between the wall
attachment cable hose outlet (A) and the system cabinet.
Figure 5-9
3
1
Table
CS
WS
SC
Control
room
2
4
A
Содержание OMNERA 400T