Wall Specifications
The walls are normally only used for mounting the cable outlets for the Ceiling suspended x-
ray support, the Wallstand and the System cabinet. Therefore, the only specification for the
walls, is that it must be possible to mount and secure the cable outlets. Screws, bolts or similar
for mounting the cable outlets, must be supplied (not included in shipment). The regulations
and specifications on page 2-1 must be followed.
The Installation cube shall be mounted to the wall. There are four wall fixation points, and each
fixation point must withstand a load of at least 0.5 kN. The wall attachment allows to place the
installation cube 0-1000 mm from the wall. If the distance to the wall is small, the attachment
must be shortened to a correct length. When the Installation cube is placed close to the wall,
make sure that a suitable placement of the cable outlet is possible.
Figure 2-7
750
750
750
750
Minimum distance, 0 mm
Maximum distance, 1000 mm
0.5kN
0.5kN
0.5kN
0.5kN
76
86
4
0
22
10.5 (2x)
Содержание OMNERA 400T
Страница 1: ...Technical Publication Planning Guide 0180 095 485 EN Rev 3 1...
Страница 4: ...Revision Planning Guide 0180 095 485 OMNERA 400T Rev 3 1...
Страница 6: ...Contents Planning Guide 0180 095 485 OMNERA 400T...
Страница 8: ...Introduction Planning Guide 0180 095 485 OMNERA 400T Page 1 2...
Страница 48: ...Anchoring Location Planning Guide 0180 095 485 OMNERA 400T Page 4 10...
Страница 68: ...Room Layout Planning Guide 0180 095 485 OMNERA 400T Page 5 20...
Страница 82: ...Component Specification Planning Guide 0180 095 485 OMNERA 400T Page 6 14...
Страница 88: ...Transport Specifications Planning Guide 0180 095 485 OMNERA 400T Page 8 2...
Страница 89: ...Intentionally left blank...
Страница 90: ...Canon CANON MEDICAL SYSTEMS USA INC...