Component Specification
OMNERA™400T
0180-095-485
Planning Guide
Page 6-5
Alpha/Beta Movements
The Alpha movement has a total stroke of 337º, limited by fixed mechanical end stops. The
Alpha movement has three mechanical indexes (-90º, 0º, 90º). The current Alpha angle is al-
ways shown in the display, placed in front of the x-ray tube.
The Beta movement has a total stroke of 343º, limited by a mechanical end stop. The mechan-
ical end stop is adjustable allowing the System to be configured to suit all possible room lay-
outs. The Beta movement has four mechanical indexes (-180º, -90º, 0º, 90º).
Figure 6-3 Movement range Beta/Alpha.
0
-90
90
180
Alpha
Beta
0
-180
-90
90
-180(-5)
+163
+180(+5)
-157