![Canon OCT-A1 Скачать руководство пользователя страница 246](http://html1.mh-extra.com/html/canon/oct-a1/oct-a1_operation-manual_4014345246.webp)
246
9 Maintenance
CAUTION
•
For safety reasons, before using the OCT, be sure to perform the daily
inspection.
•
Have a periodically inspection performed for the OCT at least once a year
by a Canon designated representative to maintain its performance and
reliability.
Daily Inspections
Perform the following inspections before using the OCT to ensure that it is used safely
and correctly. If a problem is found during the inspection and you are unable to correct
the problem, please contact your sales representative or local Canon dealer.
Checks Before Turning On the Power
Check the following items before turning on the power.
1) The power cord and connection cable are not damaged and their insulation is not
torn.
2) The power cord is fully and securely inserted into the AC connector on the OCT and
the AC outlet.
3) The exterior of the OCT is not damaged or loose.
4) There are no scratches or dirt on the objective lens. Clean the objective lens if it is
dirty (see page 247).
5) There is no dirt on the computer monitor. Clean the computer monitor if it is dirty.
Checks After Turning On the Power
Turn on the OCT, and then turn on the computer. Check the following items after you log
in.
1) The POWER lamp lights.
2) The chin rest moves up and down smoothly as the [Chin Rest] button is clicked.
3) The measurement unit moves up and down smoothly as the anterior segment image
is clicked.
4) The measurement unit moves back and forth smoothly as the [Z Adjustment] button
is clicked.
5) The SLO live image can be observed as [Start] is clicked.
BT8-1794-E01_RX OCT-A1_HS100 V4.6_CVS.indb 246
10/26/2020 10:50:10 AM
Содержание OCT-A1
Страница 9: ...1 Introduction BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 9 10 26 2020 10 49 20 AM ...
Страница 21: ...2 Safety BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 21 10 26 2020 10 49 21 AM ...
Страница 37: ...3 Part Names BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 37 10 26 2020 10 49 23 AM ...
Страница 46: ...46 3 Part Names BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 46 10 26 2020 10 49 27 AM ...
Страница 251: ...10 Troubleshooting BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 251 10 26 2020 10 50 11 AM ...
Страница 254: ...254 10 Troubleshooting BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 254 10 26 2020 10 50 11 AM ...
Страница 255: ...Appendix BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 255 10 26 2020 10 50 11 AM ...
Страница 272: ...272 BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 272 10 26 2020 10 50 11 AM ...
Страница 273: ...273 BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 273 10 26 2020 10 50 11 AM ...
Страница 274: ...274 BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 274 10 26 2020 10 50 11 AM ...
Страница 275: ...BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 275 10 26 2020 10 50 11 AM ...