240
8 Settings
Restoring Data
By restoring data, the state of data can be returned up to the stage of the selected
backup log.
Important
By restoring data, the data that is newer than the selected backup log is deleted.
Before restoring data, check the date of the backup log and the contents of data.
1
Select a backup log, and then click [Restore].
For example, if restoring the backup log of 5/16/2012, the status returns to 5/16/2012.
Therefore, the patient information and examination data acquired on and after 5/17/2012
are deleted.
Restoring starts.
When the screen where the progress bar is indicated disappears, restoring is finished.
BT8-1794-E01_RX OCT-A1_HS100 V4.6_CVS.indb 240
10/26/2020 10:50:09 AM
Содержание OCT-A1
Страница 9: ...1 Introduction BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 9 10 26 2020 10 49 20 AM ...
Страница 21: ...2 Safety BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 21 10 26 2020 10 49 21 AM ...
Страница 37: ...3 Part Names BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 37 10 26 2020 10 49 23 AM ...
Страница 46: ...46 3 Part Names BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 46 10 26 2020 10 49 27 AM ...
Страница 251: ...10 Troubleshooting BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 251 10 26 2020 10 50 11 AM ...
Страница 254: ...254 10 Troubleshooting BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 254 10 26 2020 10 50 11 AM ...
Страница 255: ...Appendix BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 255 10 26 2020 10 50 11 AM ...
Страница 272: ...272 BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 272 10 26 2020 10 50 11 AM ...
Страница 273: ...273 BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 273 10 26 2020 10 50 11 AM ...
Страница 274: ...274 BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 274 10 26 2020 10 50 11 AM ...
Страница 275: ...BT8 1794 E01_RX OCT A1_HS100 V4 6_CVS indb 275 10 26 2020 10 50 11 AM ...