CHAPTER 5 TROUBLESHOOTING
3. Scanning Does Not Start
Error message appears. The reading unit does not work.
Message Step Check
item Result Action
----(General)
1
Is the problem solved by
resetting the machine
power and restarting the
computer?
YES End.
1
Is the ADF cable con-
nected?
NO Connect.
2
Is the ADF PCB properly
connected?
NO Properly
connect.
Feeder is not
connected
3
Is the problem solved by
replacing the ADF PCB?
YES End.
1
Is the lock switch released?
NO
Release.
2
Is the software properly in-
stalled?
NO Properly
install.
Scanner is Locked
Calibration Error
Lamp Error
Hardware Error
3
Is the software version ap-
propriate?
NO
Install an appropriate
version.
1
Is the job properly set?
NO
Properly set.
Job-related error
messages
2
Is the system properly
working including the net-
work connections?
NO
Ask user to resume the
system.
Table 5-303
If the scanner does not resume normal
operation after taking the above listed ac-
tions, the problem may be with the com-
ponents inside the flatbed assembly such
as the scanner motor or the reading unit.
Other than the operation panel, the parts
inside the flatbed assembly are not as-
signed as service parts. Therefore, the
product should be replaced.
Scanning by the job function is impossible
if “Disable device events” is selected on
the Windows event setting screen.
Tens of seconds are required to warming
up the fluorescent lamp after the machine
power is turned on or the machine is re-
sumed from the power saving mode. This
is not a failure. By user operation, how-
ever, the user can change the time before
the power saving mode is on. Refer to
“Lamp Control” on the next page for de-
tails.