CR9000 Overview
OV-2
9031 CPU
FIBER OPTIC
LINK IN
FIBER OPTIC
LINK OUT
TLINK
MADE IN USA
FIGURE OV1-2. CR9031
CR9041 A/D and Amplifier Module
The CR9041 A/D and Amplifier Module provides signal conditioning and 16
bit, 100 kHz A/D conversions.
9041 A D
CR9000
MEASUREMENT & CONTROL SYSTEM
LOGAN, UTAH
MADE IN USA
FIGURE OV1-3. CR9041
CR9011 Power Supply Module and AC Adapter
The CR9011 Power Supply Module provides regulated power to the CR9000
from the internal battery modules. It also regulates battery charging from
power supplied by the AC adapter, a DC input, or other external sources. The
AC adapter may be used where AC power is available (100 - 240 volts) to
provide power to the CR9000 and charge its batteries.
The CR9011 has a relay that allows shutting off power under program control.
The Power Up inputs allow an external signal to awaken the CR9000 from a
powered down state (PowerOff, Section 9). When the CR9000 is in this power
off state the ON Off switch is in the on position but the internal relay is open.
The power LED is not lit. If the "<0.5 " input is switched to ground or if the
">2" input has a voltage greater than 2 volts applied, the CR9000 will awake,
load the program in memory and run. If the "< 0.5" input continues to be held
at ground while the CR9000 is powered on and goes through its 2–5 second
initialization sequence, the CR9000 will not run the program in memory.
MEASUREMENTS:
Battery (voltage and current)
CONTROL:
PowerOff
Содержание CR9000
Страница 6: ...CR9000 Table of Contents iv This is a blank page ...
Страница 22: ...CR9000 Overview OV 16 This is a blank page ...
Страница 26: ...Section 1 Installation 1 4 CR9000 FIGURE 1 1 3 CR9000 Battery Pack ...
Страница 72: ...Section 3 CR9000 Measurement Details 3 28 This is a blank page ...
Страница 88: ...Section 5 Program Declarations 5 6 This is a blank page ...
Страница 217: ...This is a blank page ...