![Campbell CR9000 Скачать руководство пользователя страница 114](http://html1.mh-extra.com/html/campbell/cr9000/cr9000_instruction-manual_489197114.webp)
Section 6. Data Table Declarations and Output Processing Instructions
6-26
Totalize (Reps, Source, DataType, DisableVar)
This instruction stores the total(s) of the values of the source(s) over the given
output interval.
Parameter
& Data Type
Enter
Totalize Parameters
Reps
Constant
The number of totals to calculate. When repetitions are greater than 1, the source must be an array.
Source
Variable
The name of the Variable that is the input for the instruction.
DataType
A code to select the data storage format.
Constant
Alpha Code
Numeric Code
Data Format
IEEE4
24
IEEE 4 byte floating point
FP2
7
Campbell Scientific 2 byte floating point
DisableVar
Constant,
Variable, or
Expression
A non-zero value will disable intermediate processing. Normally 0 is entered so all inputs are processed.
For example, when the disable variable is
≠
0 the current input is not included in the total. The total that
is eventually stored is the total of the inputs that occurred while the disable variable was 0.
Value Result
0
Process current input
≠
0
Do not process current input
Содержание CR9000
Страница 6: ...CR9000 Table of Contents iv This is a blank page ...
Страница 22: ...CR9000 Overview OV 16 This is a blank page ...
Страница 26: ...Section 1 Installation 1 4 CR9000 FIGURE 1 1 3 CR9000 Battery Pack ...
Страница 72: ...Section 3 CR9000 Measurement Details 3 28 This is a blank page ...
Страница 88: ...Section 5 Program Declarations 5 6 This is a blank page ...
Страница 217: ...This is a blank page ...