Section 7. Measurement Instructions
7-34
Parameter
Enter
FFTFILT PARAMETERS
Dest
Variable or
Array
The Variable in which to store the results of the instruction.
Because
FFTFilt
returns all or part of
an entire spectrum (see
ILow
and
IHigh)
for each
Rep
,
Dest
usually must be an array.
Reps
Constant
The number of times to repeat the measurements and subsequent FFTs on consecutive
CR9052DC channels. Spectra from multiple
Reps
are placed head-to-tail in the
Dest
array.
Range
Constant
The voltage range for the measurement.
The CR9052 normally replaces out-of-range
measurements with not-a-number (NaN) which is displayed in PC9000 as
Range?
. Users may
choose to have out-of-range measurements to be replaced by the analog-to-digital converter
saturation value with a special code in
FiltOption
.
Alpha Code
Numeric Code
Voltage
Range
Module Excitation
Board Supported
mV1000
0
± 5000 mV
CR9052DC, CR9052IEPE
mV1000
1
± 1000 mV
CR9052DC, CR9052IEPE
mV200
4
± 200 mV
CR9052DC
mV50
5
± 50 mV
CR9052DC
mV20
6
± 20 mV
CR9052DC
Fslot Const
The number of the slot that holds the CR9052 Module to be used for the measurement.
Chan
Constant
The CR9052 channel number on which to make the first measurement. When
Reps
are used, subsequent
measurements will be automatically made on the following differential channels.
FiltOption
Constant
The sample ratio for the measurement (how many measurements are made within one cycle of
the highest frequency in the pass band). The sample ratio determines the top of the pass-band
(
F
PASS
) and the beginning of the stop-band (
F
STOP
) of the anti-aliasing low-pass filter relative
the sample rate (
F
SAMPLE
). The sample rate is the inverse the scan interval in the CRBASIC
program.
FiltOption
must be the same for all channels of a single CR9052DC Filter Module.
The CR9052 normally replaces out-of-range measurements with not-a-number (NaN) which is
displayed in PC9000 as
Range?
. Out-of-range measurements may be replaced by the analog-
to-digital converter saturation value by adding 1000 to the
FiltOption
codes shown below.
Numeric Code
Sampling Ratio
F
PASS
F
STOP
2 2.5
F
SAMPLE
/2.5
F
SAMPLE
/2.01
5 5
F
SAMPLE
/5
F
SAMPLE
/3.37
10 10
F
SAMPLE
/10
F
SAMPLE
/5.08
20 20
F
SAMPLE
/20
F
SAMPLE
/6.81
1* 2.155
23.2 Khz
26.8 kHz
FiltOption
1 is available only when
FsampRate
is 50 kHz. At this sample rate, no additional
filtering beyond that provided by the CR9052DC hardware is required to anti-alias the data.
Because the CR9052DC processor is not performing additional anti-alias filtering, this
FiltOption
increases the CR9052DC’s FFT throughput. To achieve spectra from seamless
snapshots with
FsampRate
equal to 50 kHz on six channels,
FiltOption
must be 1.
Содержание CR9000
Страница 6: ...CR9000 Table of Contents iv This is a blank page ...
Страница 22: ...CR9000 Overview OV 16 This is a blank page ...
Страница 26: ...Section 1 Installation 1 4 CR9000 FIGURE 1 1 3 CR9000 Battery Pack ...
Страница 72: ...Section 3 CR9000 Measurement Details 3 28 This is a blank page ...
Страница 88: ...Section 5 Program Declarations 5 6 This is a blank page ...
Страница 217: ...This is a blank page ...