SECTION 13. CR7 MEASUREMENTS
13-8
Table 13.3-6 summarizes maximum lead
lengths for corresponding error limits in six
Campbell Scientific sensors. Since the first
three sensors are non-linear, the voltage error,
Ve, is the most conservative value
corresponding to the error over the range
shown.
MINIMIZING SETTLING ERRORS IN NON-
CAMPBELL SCIENTIFIC SENSORS
When long lead lengths are mandatory in
sensors configured by the user, the following
general practices can be used to minimize or
measure settling errors:
1.
When measurement speed is not a prime
consideration, Instruction 4 (Excite, Delay
and Measure) can be used to ensure ample
settling time for half-bridge, single-ended
sensors.
2.
An additional low value bridge resistor can
be added to decrease the source
resistance, Ro. For example, assume a YSI
non-linear thermistor such as the model
44032 is used with a 30 kohm bridge
resistor, R'f. A typical configuration is
shown in Figure 13.3-7A. The
disadvantage with this configuration is the
high source resistance shown in column 3
of Table 13.3-7. Adding another 1K
resistor, Rf, as shown in Figure 13.3-7B
lowers the source resistance of the CR7
input but offers no improvement over
configuration A because R'f still combines
with the lead capacitance to slow the signal
response at point P. The source resistance
at point P (column 5) is essentially the
same as the input source resistance of
configuration A. Moving Rf' out to the
thermistor as shown in Figure 13.3-7C
optimizes the signal settling time because it
becomes a function of Rf and Cw only.
Columns 4 and 7 list the signal voltages as a
function of temperature with a 5V excitation for
configurations A and C, respectively. Although
configuration A has a higher output signal (5V
input range), it does not yield any higher
resolution than configuration C which uses the
±150 mV input range.
NOTE:
Since Rf' attenuates the signal in
configurations B and C, one might consider
eliminating it altogether. However, its
inclusion "flattens" the non-linearity of the
thermistor, allowing more accurate curve
fitting over a broader temperature range.
TABLE 13.3-6. Maximum Lead Length vs. Error for Campbell Scientific Resistive Sensors
Sensor
Maximum
Model #
Error
Range
Ve(µV)
Length(ft.)
107
0.05oC
0oC to 40oC
5
9651
207(RH)
1%RH
20% to 90%
500
19503
WVU-7
0.05oC
0oC to 40oC
5
8502
024A
3o
@ 360o
1390
2502
227
-
-
-
20003
237
10 kohm
20k to 300k
500
19003
1 based on transient settling
2 based on signal rise time
3 limit of excitation drive
Содержание CR7
Страница 2: ...This is a blank page ...
Страница 4: ...This is a blank page ...
Страница 28: ...CR7 MEASUREMENT AND CONTROL SYSTEM OVERVIEW OV 18 This is a blank page ...
Страница 38: ...SECTION 1 FUNCTIONAL MODES 1 10 This is a blank page ...
Страница 42: ...SECTION 2 INTERNAL DATA STORAGE 2 4 This is a blank page ...
Страница 52: ...SECTION 3 INSTRUCTION SET BASICS 3 10 This is a blank page ...
Страница 57: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 5 FIGURE 4 4 1 Example of CR7 Printable ASCII Output Format ...
Страница 58: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 6 This is a blank page ...
Страница 68: ...SECTION 6 9 PIN SERIAL INPUT OUTPUT 6 6 This is a blank page ...
Страница 140: ...SECTION 13 CR7 MEASUREMENTS 13 16 FIGURE 13 5 1 Circuits Used with Instructions 4 9 ...
Страница 162: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Страница 164: ...B 2 This is a blank page ...
Страница 170: ...APPENDIX C BINARY TELECOMMUNICATIONS C 6 This is a blank page ...
Страница 173: ...APPENDIX D CALIBRATION PROCEDURES D 3 FIGURE D 2 1 Calibration Points for the Analog Interface Card ...
Страница 174: ...APPENDIX D CALIBRATION PROCEDURES D 4 FIGURE D 2 2 CR7X CPU Card ...
Страница 186: ...CR7 INDEX I 8 This is a blank page ...