
SECTION 13. CR7 MEASUREMENTS
13-7
EXAMPLE LEAD LENGTH CALCULATION
FOR CAMPBELL SCIENTIFIC 107
TEMPERATURE SENSOR
Assume a limit of 0.05oC over a 0oC to +40oC
range is established for the transient settling
error. This limit is a reasonable choice since it
approximates the linearization error over that
range. The output signal from the thermistor
bridge varies non-linearly with temperature
(refer to 107 Operator's Manual), ranging from
about 200 µV/oC at 0oC to 100 µV/oC at 40oC.
Taking the most conservative figure yields an
error limit of Ve = 5 µV. The other values
needed to calculate the maximum lead length
are summarized in Table 13.3-5 and listed
below:
1) Veo
≈
100mV, peak transient at 4V excitation
2) Ve
≈
5µV, allowable measurement error
3) t = 500µs, CR7 input settling time
4) Ro = 1kohm, 107 probe source resistance
5) Cf = 0.01nfd, CR7 input capacitance
6) Cw = 42pfd/ft., lead wire capacitance
Solving Equation 13.3-15 gives a maximum
lead length of:
L
≈
965 ft., error
≈
0.05oC
Setting the allowable error at 0.1oC or
approximately 10µV, the maximum lead length
increases to:
L
≈
1050 ft., error
≈
0.1oC
13.3.4 SUMMARY OF SETTLING ERRORS FOR
CAMPBELL SCIENTIFIC RESISTIVE
SENSORS.
Table 13.3-5 summarizes the data required to
estimate the effect of lead length on settling
errors for Campbell Scientific's resistive
sensors. Comparing the transient level, Veo, to
the input range, one suspects that transient
errors are the most likely limitation for the 107
sensor. The sensors in the WVU-7 are the
same as in the Model 107 (the lead wire is
different) but the signal leads for the WVU-7 wet
and dry bulbs are not subject to excitation
transients because they are shielded
independently from the excitation.
The comparatively small transient yet large
source resistance of the 024A sensor indicates
that signal rise time may be the most important
limitation. The analysis in Section 13.3.2
confirms this.
The Model 227 Soil Moisture Block has a
relatively short time constant and essentially no
transient. Lead lengths in excess of 2000 feet
produce less than a 0.1 bar (0-10 bar range)
input settling error. With this sensor, the drive
capability of the excitation channel limits the
lead length. If the capacitive load exceeds 0.1
ufd and the resistive load is negligible, Vx will
oscillate about it's control point. If the capacitive
load is 0.1 ufd or less, Vx will settle to within
0.1% of its correct value in 150µs. A lead
length of 2000 feet is permitted for the Model
227 before approaching the drive limitation.
TABLE 13.3-5. Summary of Input Settling Data for Campbell Scientific Resistive Sensors
Sensor
Belden
R
o
C
w
τ
*
Input
Model #
Wire #
(kohms) (pfd/ft.) (us)
Range(mV)
V
x
(mV)
V
eo
(mV)**
107
8641
1
42
52
15
4000
100
207(RH)
8771
1
41
51
150
3000
130
WVU-7
8723
1
62
72
15
4000
0
227
8641
0.1-1
42
5-52
500
500
0
237
8641
1
42
52
50
5000
125
024A
8771
0-5
41
1-255
500
1000
0-90
*
Estimated time constants are for 1000 foot lead lengths and include 0.01nfd CR7 input capacitance.
**
Measured peak transients for 1000 foot lead lengths at corresponding excitation, Vx.
Содержание CR7
Страница 2: ...This is a blank page ...
Страница 4: ...This is a blank page ...
Страница 28: ...CR7 MEASUREMENT AND CONTROL SYSTEM OVERVIEW OV 18 This is a blank page ...
Страница 38: ...SECTION 1 FUNCTIONAL MODES 1 10 This is a blank page ...
Страница 42: ...SECTION 2 INTERNAL DATA STORAGE 2 4 This is a blank page ...
Страница 52: ...SECTION 3 INSTRUCTION SET BASICS 3 10 This is a blank page ...
Страница 57: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 5 FIGURE 4 4 1 Example of CR7 Printable ASCII Output Format ...
Страница 58: ...SECTION 4 EXTERNAL STORAGE PERIPHERALS 4 6 This is a blank page ...
Страница 68: ...SECTION 6 9 PIN SERIAL INPUT OUTPUT 6 6 This is a blank page ...
Страница 140: ...SECTION 13 CR7 MEASUREMENTS 13 16 FIGURE 13 5 1 Circuits Used with Instructions 4 9 ...
Страница 162: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Страница 164: ...B 2 This is a blank page ...
Страница 170: ...APPENDIX C BINARY TELECOMMUNICATIONS C 6 This is a blank page ...
Страница 173: ...APPENDIX D CALIBRATION PROCEDURES D 3 FIGURE D 2 1 Calibration Points for the Analog Interface Card ...
Страница 174: ...APPENDIX D CALIBRATION PROCEDURES D 4 FIGURE D 2 2 CR7X CPU Card ...
Страница 186: ...CR7 INDEX I 8 This is a blank page ...